FlowCodeUnmix is intended to be a complete debarcoding and unmixing pipeline for spectral flow cytometry samples using the FlowCode protein epitope barcoding technology.
FlowCodeUnmix is provided under an AGPL3 licence.
To read about FlowCodes see the Immunity paper by Bricard et al. and also the ProCode system upon which FlowCodes are based.
You will need to install the following packages:
- BiocManager
- FlowSOM
- flowWorkspace
- remotes (or devtools)
- AutoSpectral
Everything else should be installed automatically.
Run the code below on your machine to set everything up.
# Install Bioconductor packages
if (!requireNamespace("BiocManager", quietly = TRUE)) install.packages("BiocManager")
BiocManager::install(c("flowWorkspace", "flowCore", "FlowSOM", "remotes"))
# You'll need devtools or remotes to install from GitHub.
if (!requireNamespace("pak", quietly = TRUE)) install.packages("pak")
if (!requireNamespace("remotes", quietly = TRUE)) install.packages("remotes")
remotes::install_github("DrCytometer/AutoSpectral")
remotes::install_github("DrCytometer/FlowCodeUnmix")FlowCodeUnmix is written in R and should work on any system. R, however, is not very fast, so the per-cell unmixing part will be slow. To speed this up, install the Rcpp version available at FlowCodeUnmixRcpp. This is approximately 100x faster.
You will need these pieces of information:
- Single-stained control samples, preferably cell-based
- A “backbone”” control sample stained with all the anti-epitope antibodies (and nothing else)
- Unstained control cells matching the source used in the backbone control
- Unstained control cells matching the source(s) in your fully stained samples
- A CSV file describing the valid combinations of epitope barcodes and what they correspond to. Example format below:
FlowCodeUnmix is based around AutoSpectral. Please see the help pages and articles for AutoSpectral as a starting point. In particular, see the AutoSpectral Full Workflow.
For FlowCodeUnmix, see the example workflow in “FlowCode_Unmixing_Workflow.Rmd”, which is also available on the web at FlowCode Workflow. You can view all of the functions available in FlowCodeUnmix here.
Here is what the unmixing looks like for the backbone control (FlowCode
epitopes only) with standard OLS unmixing or using FlowCodeUnmix:

You get debarcoded channels in the output FCS file: 
The current implementation (unmix.flowcode.cpp.staged()) runs a
five-stage pipeline that brackets a thin FlowCode-specific correction
between two calls into the generic, FlowCode-agnostic AutoSpectral joint
unmixing engine:
- Initial unmix. A first, generic AutoSpectral joint unmix
(
AutoSpectralRcpp::unmix.autospectral.rcpp()) is run on the raw data. Off-target precision doesn’t matter yet here — this pass only needs to produce fluorophore coefficients clean enough to debarcode on. - Debarcode.
debarcode()assigns each cell a FlowCode triplet combination ID (or “untransduced”/“unrecognised”) from the stage 1 coefficients, using the per-fluorophore positivity thresholds. - Pooled cross-combo background subtraction. For each FlowCode fluorophore, we pool the stage 1 coefficient values across every valid-ID cell whose combo does not have that fluorophore active, and take one population-wide median. This gives a single non-specific-background estimate per FlowCode fluorophore, which is reconstructed in raw/spectral space and subtracted uniformly from every cell — the correction doesn’t vary by combo identity or debarcode ID. Pooling across combos (rather than per-combo) is what separates true background from FRET: any one combo’s systematic FRET in a channel gets diluted out by every other combo where that channel isn’t elevated.
- FRET fit. For valid-ID cells only, every candidate FRET spectrum for the cell’s combo is fit jointly, in a small OLS, together with the combo’s own active tag fluorophores and the fixed-shape mean AF spectrum, against the background-corrected raw data. The candidate minimising the L1 residual is kept, and its fitted contribution is subtracted from that cell’s raw data. Off-target FlowCode channels are already handled by the background step, so they aren’t part of this fit’s basis at all.
- Final unmix. A second, generic AutoSpectral joint unmix is run on the background-and-FRET-corrected raw data. This pass is the one that matters for accuracy: for each cell, it works in the reduced space of fluorophores that are actually positive (plus AF, which is always considered present), testing candidate spectral variants for each positive fluorophore and for AF and keeping whichever minimises the residual. This is the reported result.
- FlowCode and tag channels are built from the final unmixed
coefficients using the debarcoded combination IDs, giving a
FlowCodetransduction channel and one channel per valid combination.