backend/hardware-controller: The flowrate has an incorrect calibration #289

@pkostyrka

Description

@pkostyrka

Describe the bug
With a 2045 step/ml for the pump with a flowrate of 5 mL/min, when you pass 10 mL, the pump effectively pass 10 mL.
But, it takes around 3 minutes, when it should be 2 minutes. So the real flowrate is around 3.5 mL/min.
(I did this on ten units).

To Reproduce
Steps to reproduce the behavior:

  1. OPTIC CONFIGURATION
  2. Pass 10 mL with a flowrate of 5 mL/min. Measure the time. It should be 2 minutes if the real flowrate is 5 mL/min. But it is not the case.

Expected behavior