Reverse-engineered from CyberView X5 5.16.75 (plugins/DD6000.dll, the driver plugin for
USB ID 05e3:0142), and verified against USB captures of CyberView driving a
Reflecta DigitDia 4000 (INQUIRY model 0x2f, firmware 1.14) and against the hardware.
The SANE pieusb backend documents the same command family for related PIE scanners.
The scanner is a parallel-port SCSI design behind a Genesys USB bridge
(one vendor interface: bulk EP 0x81 in, bulk EP 0x02 out, interrupt EP 0x83).
CyberView talks to it through usbscan.sys IOCTL_WRITE_REGISTERS (0x80002010) /
IOCTL_READ_REGISTERS (0x8000200c), which become vendor control transfers:
| Transfer | bmRequestType | bRequest | wValue | Data |
|---|---|---|---|---|
| write register | 0x40 |
0x0C |
register | 1 byte |
| read register | 0xC0 |
0x0C |
register | 1 byte |
| bulk header | 0x40 |
0x04 |
0x82 |
[dir,0,0,0,len LE32] (dir 0 = in, 1 = out) |
Registers: 0x88 IEEE-1284 data, 0x87 IEEE-1284 control, 0x85 SCSI byte FIFO,
0x84 phase/status, 0x82 bulk size.
IEEE-1284 select (FUN_1002e3c0): write ff aa 55 00 ff 87 78 <cmd> to 0x88,
3 ms pause, 05, 04 to 0x87, ff to 0x88. cmd: 0xE0 = SCSI command,
0x30 = reset, 0x00 = address.
SCSI command (FUN_10030d80):
- select
0xE0, write the 6 CDB bytes to0x85, read the phase byte from0x84; - phase
1(data in): bulk header with the length, read bulk EP0x81(chunks ≤0xFFF0); phase0(data out): write each byte to0x85(< 256 bytes) or bulk header withdir=1and write EP0x02; then read0x84again; - phase
3= done; the next read of0x84is the SCSI status:0good,2check condition (issue REQUEST SENSE),8busy (retry). - any other phase: recovery = write
04to0x87, select0x30.
Connect sequence (capture): 0x87←04, select 0x30 ×2, select 0x00, INQUIRY,
select 0x30/0x00, bridge check (0x8c←84 with wIndex 0x10, read 0x8e, twice).
Timing. While motors run, the bridge NAKs control transfers for several seconds. CyberView waits patiently (one TEST UNIT READY in the capture blocked 4.9 s). Hosts that abort a control transfer pending for ~5 s (e.g. some VM USB passthrough) wedge the bridge until power-cycle, so the driver stays off the bus for 15 s after SCAN.
| Op | Name | Notes |
|---|---|---|
00 |
TEST UNIT READY | 02/04/01 while the lamp warms up |
03 |
REQUEST SENSE | 14 bytes |
08 |
READ | count = lines (image/shading) or bytes (parameter blocks) |
0A |
WRITE | parameter blocks, see below |
0F |
PARAM | 18 bytes: width, lines, bytes/line (LE16) — after SCAN only |
12 |
INQUIRY | 0xA4 bytes; model at 116 (0x2f = DigitDia 4000) |
15 |
MODE SELECT | 16 bytes, see below |
18 |
COPY (CCD mask) | width bytes; 0 marks CCD elements used at this resolution |
1A |
MODE SENSE | 16 bytes |
1B |
SCAN | len 1 = start, 0 = stop |
D1 |
SLIDE | 4 bytes [action, count/focus LE16, 0] |
D7 |
READ GAIN/OFFSET | 103 bytes |
DC |
WRITE GAIN/OFFSET | 29 bytes: exposure R,G,B (LE16), offset R,G,B @6, gain R,G,B @12, lamp level @15, IR exposure @18, IR offset @20, IR gain @22 |
DD |
READ STATE | 12 bytes: [2] magazine position (0-based), [3] focus, [5] busy, [6]&7==5 magazine ready, [6]&0x40 scanning, [8] lamp in power-save |
F2 |
LAMP ON | no data; wakes the lamp from power-save. The answer only comes after the lamp has ignited (> 5 s) |
WRITE blocks ([code LE16, len LE16, …]): 0x12 scan frame [index=0x80, x0, y0, x1, y1]
in optical (3600 dpi) units; 0x13 exposure [filter, percent]; 0x14 highlight/shadow
[filter, percent] (filters 2/4/8 = R/G/B); 0x95 prepare reading the shading
parameters (then READ 0x80: lines @11, width @12); 0x17 CMD_17 (rejected by the
DigitDia 4000 with 05/26/80; CyberView ignores that).
MODE SELECT: [0, 0x0f, res LE16, passes, depth, format=4, 0, order=1, quality, 0, 0, final, 0x80, 0x10, 0];
passes 0x80 RGB, 0x90 RGB + infrared; depth 0x04 8 bit, 0x20 16 bit;
quality 0x08 = skip shading collection; final = 1 for the final (non-preview) scan.
SLIDE actions on the DigitDia 4000: 0x10 load the slide at the current magazine position
into the scanner (byte 1 = focus; fixed 0x73 on this model) — SCAN must follow within
about a second or the slide is returned; 0x04/0x05 move the magazine forward/back
by count; 0x40 reload the magazine.
- exposure 100 % and highlight/shadow 100 % for R, G, B; scan frame; CMD_17;
- READ GAIN/OFFSET, WRITE GAIN/OFFSET; MODE SELECT;
- SLIDE
[10 73 00 00], 0.5 s, TEST UNIT READY, SCAN; - on the first scan of a session: D7/DC, READ 4 shading lines, READ the rest minus 4,
D7/DC, READ 4 (40 lines × R, G, B, I at
2·width+2bytes, 16 bit, taggedRR/GG/BB/II); - CCD mask, PARAM, then READ image lines in blocks of ≤ 255; each line starts with a
two-byte colour tag (
RR,GG,BB,II); colours arrive interleaved per line; - READ STATE until not busy (slide returned to the magazine).
Between slides CyberView sends SLIDE [04 01 …] and polls READ STATE until the position
increments. Per slide it previews at 300 dpi with fixed exposures (R/G/B 0x7ac2/0x7582/0x4daf,
gain 0x19), derives exposures from the preview, and scans the final image with gains
0x36/0x3c/0x3f and correspondingly shorter exposures (exposure × gain is preserved;
gain curve from pieusb).
After about 15 minutes without a scan the scanner switches the lamp off and sets READ STATE
byte 8 (polling does not prevent this). Loading a slide in that state fails: SCAN answers
02/04/01, the lamp needs about a minute, but the transport gives up on the waiting slide
after ~25 s and latches 04/81/82. CyberView (FUN_10019320) therefore checks byte 8 before
each scan and, if set, waits 2 s, sends LAMP ON (F2) and polls READ STATE every ~3 s
until byte 5 (busy) clears, before loading the slide.
| Sense | Meaning |
|---|---|
02/04/01 |
lamp warming up |
05/20/00 |
command not valid now (e.g. READ STATE during a scan) |
05/26/80 |
not supported by this model (CMD_17) |
09/30/01 |
no slide in the scan position |
04/81/82 |
slide transport fault (magazine or slide lift disturbed) — power-cycle |