@jkabisch

Hi,

thanks for the excellent PocketPCR which we apply both in the lab and for teaching. We stumbled upon a bug which we here would like to report:

Expected behavior:
After a cycling program has been finished, a new program can be started with "Run PCR"

Actual behavior:
The program starts, but it does not cycle.

Workaround:
Unplug the PocketPCR from the power, plug in again and now running a program and cycling works as expected.